Detail of Recall

Type of Product:Device
Medsafe Reference:13715
Brand Name:Roche cobas c702 Module
Model:c702
Affected:All versions
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer advises there is potential for a mismatch between reagent probe positioning on the cobas c 702 and the aspirated reagent volumes to occur in rare instances and may affect patient and control recovery.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/10/2012