Detail of Recall

Type of Product:Device
Medsafe Reference:13698
Brand Name:Hospira Plum A+ Pump Version 11.6
Model:611239101
Affected:12391, 111971
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Brand Name:Hospira Plum A+ Pump Version 13.4 Med net ABG DriverInfusion pump
Model:612079201
Affected:20792
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:The volume control knob located on the back of the infuser may not function as described in the System Operating Manual; the direction for loud and quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume, however on some devices the alarm volume increases when the knob is turned clockwise. While the device will continue to alarm with an audible tone of at least 50dBA at the lowest level, inadvertently setting a low one may result in an audible alarm that is more difficult to hear.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2012