Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13698 |
| Brand Name: | Hospira Plum A+ Pump Version 11.6 |
| Model: | 611239101 |
| Affected: | 12391, 111971 |
| Software version: | |
| Recalling Organisation: | Hospira NZ Ltd 58 Richard Pearse Drive Airport Oaks Mangere Auckland 2022 |
| Contact Information: | 04 384 7463 |
| Manufacturer: | Hospira Inc |
| Brand Name: | Hospira Plum A+ Pump Version 13.4 Med net ABG DriverInfusion pump |
| Model: | 612079201 |
| Affected: | 20792 |
| Software version: | |
| Recalling Organisation: | Hospira NZ Ltd 58 Richard Pearse Drive Airport Oaks Mangere Auckland 2022 |
| Contact Information: | 04 384 7463 |
| Manufacturer: | Hospira Inc |
| Issue: | The volume control knob located on the back of the infuser may not function as described in the System Operating Manual; the direction for loud and quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume, however on some devices the alarm volume increases when the knob is turned clockwise. While the device will continue to alarm with an audible tone of at least 50dBA at the lowest level, inadvertently setting a low one may result in an audible alarm that is more difficult to hear. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/11/2012 |