Type of Product:Device
Medsafe Reference:13695
Brand Name:Philips MammoDiagnost DR X-Ray system with advanced stereo
Model:
Affected:All units with advanced stereo option or upgrade
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:Philips advises that the stereo clutch of the Gantry might get misadjusted or its bearing might break during operation. This might cause the object table to move, resulting in a small potential risk that the table may move with the tube arm in a situation when this should not happen, e.g. when the breast is compressed and a biopsy is performed.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/10/2012