Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13695 |
| Brand Name: | Philips MammoDiagnost DR X-Ray system with advanced stereo |
| Model: | |
| Affected: | All units with advanced stereo option or upgrade |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | Philips advises that the stereo clutch of the Gantry might get misadjusted or its bearing might break during operation. This might cause the object table to move, resulting in a small potential risk that the table may move with the tube arm in a situation when this should not happen, e.g. when the breast is compressed and a biopsy is performed. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/10/2012 |