Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13680 |
| Brand Name: | GE Giraffe & Panda Infant Warmers with Resuscitation Systems with blender |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Ohmeda Medical Inc |
| Issue: | The manufacturer advises there is a potential safety issue wherein the Oxygen and Air wall inlet fittings and/or labels on the back panel, were reversed during assembly. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/10/2012 |