Detail of Recall

Type of Product:Device
Medsafe Reference:13674
Brand Name:Biomerieux BacT/ALERT System
Model:411814 or 411818
Affected:Firmware version B.40.xx
Software version:
Recalling Organisation:BioMerieux New Zealand Ltd
Suite 4, The Business Connection Office
2 Kalmia Street
Auckland 1544
Contact Information:0800 284 825
Manufacturer:BioMerieux Inc
Issue:BacT/ALERT instrument software architecture is linearly designed to scan, proces and load one bottle at a time.
Manufacturer recently launched the new BacT/ALERT 3D B.40.xx firmware that contains additional features such as bottle recalculation and bottle authentication. As a result of the additional features customers with Control Modules containing LBC-586 CPUs, distributed from 1997 through November 2007, and running B.40 firmware may exhibit a slightly slower instrument interface issue that may result in bottles being improperly loaded into the instrument.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/10/2012