Detail of Recall

Type of Product:Device
Medsafe Reference:13641
Brand Name:Nihon Kohden Automatic External Defibrillator
Model:Models TEC-5521K & TEC-5531K
Affected:
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Nihon Kohden Corporation
Issue:When the defibrillator operates on battery and the remaining battery power is zero, the battery mark becomes "0", the charge battery message appears and an alarm sounds. If this occurs and the battery isn't replaced with charged battery or the defibrillator is not connected to AC power, the mother board might become damaged and the defibrillator might not work.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Retail
Date Commenced:18/10/2012