Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13622 |
| Brand Name: | Philips DigitalDiagnost, patient carriage |
| Model: | |
| Affected: | Serial number range 00-000-00 to 09-000-543 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | The manufacturer advises the footplate of the patient carriage can be fixed in a vertical position with a hook. The hook may fail when moving the carriage, or when the operator does not secure it correctly upon moving the plate up, and the footplate can fall down. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/10/2012 |