Detail of Recall

Type of Product:Device
Medsafe Reference:13622
Brand Name:Philips DigitalDiagnost, patient carriage
Model:
Affected:Serial number range 00-000-00 to 09-000-543
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer advises the footplate of the patient carriage can be fixed in a vertical position with a hook. The hook may fail when moving the carriage, or when the operator does not secure it correctly upon moving the plate up, and the footplate can fall down.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/10/2012