Detail of Recall

Type of Product:Device
Medsafe Reference:13621
Brand Name:Medi-vac Flex Advantage Suction Liners and Canisters
Model:65652-6ll, 65652-616, 65652-631, 65651-910C, 65651-920C and 65651-930C
Affected:All lots
Software version:
Recalling Organisation:Medline International New Zealand Limited
14 George Bourke Drive
Mt Wellington
Auckland 1060
Contact Information:
Manufacturer:Cardinal Health
Issue:Safety Alert Notice advising customers that some users of the product:
1. May be testing functionality by "bumping" the Medi-Vac liners. As indicated in the Warning section of the product's DFU, premature shut-off of the vacuum may occur if the canister is bumped.
2. May be testing vacuum/ suction by occluding the Medi-Vac liner's patient port with their finger. This practice may activate the mechanical shut-off and consequently turn offf the vacuum. Occlusion of the patient port as a testing method is not included in the product's DFU and therefore is not a recommended test.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/10/2012