Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13619
Brand Name:Animas 2020 insulin pump
Model:2020
Affected:All serial numbers
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Animas Corporation
Brand Name:Animas IR1200 Insulin Pump
Model:Model: IR1200
Affected:All serial numbers
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Animas Corporation
Issue:The manufacturer advises that the pump will cease to operate when the date changes from 31-Dec-2015 to 01-Jan-2016
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Public Recall
Date Commenced:21/12/2012