Type of Product:Device
Medsafe Reference:13600
Brand Name:Philips Digital Diagnost, mobile detector holder
Model:System codes: 712020, 712022, 712025
Affected:Single Detector, Dual Detector, and Release 3
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The Wireless Portable Detector (WPD) is fixed in the Mobile Detector Holder via a lock mechanism. The 4 countersunk screws, that are used to attach the lock mechanism, can become loose. If this happens and the mechanism fails, the detector may fall down when the holder is rotated 90° clockwise.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory