Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13592 |
| Brand Name: | Stryker Surgical Light, spring arm |
| Model: | Catalogue Numbers: 0682001270, 0682001274, 0682000226, 0682400059I |
| Affected: | Selected lots |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Communications Inc |
| Issue: | Stryker has become aware that there is a small likelihood of incomplete engagement of the Spring Arm Circlip following service or replacement of the Spring Arm. This has the potential to cause the Spring Arm to fall along with the surgical lights/monitors/navigation attachment mounted to the end of the Spring Arm. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/10/2012 |