Detail of Recall

Type of Product:Device
Medsafe Reference:13572
Brand Name:Siemens Dimension Vista IRON Flex reagent cartridge
Model:Catalogue: K3085 / Siemens Material Number (SMN): 10445135
Affected:All lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed the potential for falsely elevated results with IRON Flex® reagent cartridges when tests are processed from a well set contaminated with iron from environmental sources. All lots of IRON Flex® reagent cartridges have the potential for this issue. The estimated rate is approximately 1 out of 2,500 wells.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/10/2012