Detail of Recall

Type of Product:Device
Medsafe Reference:13569
Brand Name:Abbott Architect HAVAB G assay
Model:6C29-20, 6C29-25
Affected:20 lots
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott GmbH & Co. KG
Issue:The ARCHITECT HAVAb-IgG assay shows a shift in the serum and plasma results to higher S/CO values over time, potentially leading to reduced specificity and increased false reactive results.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/10/2012