Type of Product:Device
Medsafe Reference:13561
Brand Name:Philips Heartstart XL Defibrillator/ Monitor Battery
Model:M3516A
Affected:Batteries labeled with "Made in Taiwan" and an R-2011-12 Date of Manufacture
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Manufacturer advises that if the XL Defibrilator / Monitor (M4735A) is used on a patient when AC power not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/9/2012