Detail of Recall

Type of Product:Device
Medsafe Reference:13541
Brand Name:Elekta MOSAIQ
Model:
Affected:all versions prior to 2.41
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Impac Medical Systems Inc
Issue:A patient was mistreated after a user made and saved an inadvertent change to the definition of a treatment field (Field Edit Delta Display feature) in MOSAIQ. The manufacturer has issued additional instructions for use until a software patch is available.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/9/2012