Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13525
Brand Name:Heartsine PAD300 & PAD300P Auto External Defibrillators
Model:PAD300 & PAD300O
Affected:S/N ranges 0400000501-0700032917, 08A00035000-10A0070753, 10C00200000-10C00210106
Software version:
Recalling Organisation:Amtech Medical Ltd
281 Heads Road
Wanganui 4501
Contact Information:0800 268 324
Manufacturer:HeartSine Technologies Ltd
Issue:The manufacturer advises there are two issues with these devices;
1) The device may turn itself on without input from the user, potentially depleting the battery.
2) Some versions of the battery management software may interpret a temporary drop in battery voltage as signalling a low battery and turn the device off
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Public Recall
Date Commenced:27/9/2012