Detail of Recall

Type of Product:Device
Medsafe Reference:13497
Brand Name:Siemens Flex Large 4 Coil for MAGNETOM Skyra, Verio, Spectra and Biograph mMR
Model:Part No. 8625761
Affected:All units
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Heating may occur if each of the following conditions applies at the same time:
- coil positioned off-center in head-foot direction by more than +/- 5 cm,
- coil at side position
- application of the high RF power over several minutes
In that position , the surface temperature may well exceed temperatures defined by regulatory standards and may cause local burns to the patient.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/9/2012