Type of Product:Device
Medsafe Reference:13467
Brand Name:Siemens Artiste Linear Accelerator with Syngo RT Therapist & Syngo RT Oncologist
Model:
Affected:Ver 4.2
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:The manufacturer advises that as a consequence of an incompatibility of mixes software versions an incorrect offset may be applied for patient positioning and thys may result in dose being delivered to the wrong location.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/9/2012