Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13456 |
| Brand Name: | Varian Eclipse Treatment Planning Software |
| Model: | H48 |
| Affected: | Version 8.9 & 10 |
| Software version: | |
| Recalling Organisation: | Varian Medical Systems Australasia Pty Ltd (NZ) Millenium Centre, Level 3, Building C, 600 Great South Road Ellerslie Auckland 1051 |
| Contact Information: | 0800 486 732 |
| Manufacturer: | Varian Medical Systems Inc |
| Issue: | The manufacturer has identified that an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/9/2012 |