Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13438
Brand Name:Philips Heartstart MRx Monitor/Defib
Model:M3536A
Affected:Manufactured between 17-Apr-2012 and 8-May-2012
Software version:
Recalling Organisation:Laerdal (New Zealand) Ltd
79 Apollo Drive
Rosedale
Auckland 0632
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Due to corrupt data card some HeartStart MRx defibrillator/monitors manufactured between 17-Apr-2012 and 8-May-2012 may have incomplete 12 lead program files, and if so will present the following behaviour:
1/ Device cannot analyse a 12 lead ECG report. Due to this a user cannot view, print or transmit a 12 lead ECG.
2/ A message "Cannot Analyse ECG' is displayed on the MRx when an attempt to acquire a 12 lead report is made.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/9/2012