Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13349
Brand Name:Synthes Vectra, drill bit 2.5mm
Model:03.613.010
Affected:All lots
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Brand Name:Synthes Vectra, adjustable drill stop
Model:03.613.012
Affected:All lots
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Brand Name:Synthes Vectra, 2.5mm drill bit, adjustable depth
Model:03-613-011
Affected:All lots
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Issue:The manufacturer advises that there is the potential for the drill stop to slip on the drill bit resulting in an adverse event. If the drill stop slips on the drill bit, it could result in the penetration of the posterior cortex, dural tear requiring repair, damage to the spinal cord, quadraplegia, and the risk of a potentially life-threatening event.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/8/2012