Detail of Recall

Type of Product:Device
Medsafe Reference:13325
Brand Name:Lima SMR System glenoid components
Model:all models
Affected:all batch numbers
Software version:
Recalling Organisation:Enovis Surgical NZ Ltd
46 Hillside Road
Wairau Valley
AUCKLAND 0627
Contact Information:09 3606010
Manufacturer:Limacorporate SpA
Issue:Monitoring and analysis of the clinical results of the SMR L2 metal Back Glenoid Component indicate that under certain conditions, for example rotator cuff failure or patient trauma, the polyethylene insert can disassociate. The total revision rate of the L2 version of the SMR Metal Back Glenoid Componet in Australia is 6.78% at three years.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:23/8/2012