Detail of Recall

Type of Product:Device
Medsafe Reference:13321
Brand Name:GE Carescape B650 Patient Monitor
Model:B650
Affected:All version numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Issue:The manufacturer advises there are several potential issues with this device;
1/ The SpO2 saturation limit alartm does not always active with the Nellcor SpO2
2/ The device may enter into a constant reboot loop causing the monitor to reboot continuously.
3/ The device may reboot when connected to the Unity ID
4/ The device may not update the software for power management card.
5/ An insecurely connected or faulty USB cable may cause the monitor to reboot.
6/ There may be printing delays with the PRN50 printer
7/ The device may reboot due to an X server freeze
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/8/2012