| Type of Product: | Device |
| Medsafe Reference: | 13319 |
| Brand Name: | GE Dash 3000/4000/5000 Patient Monitor |
| Model: | |
| Affected: | Ver 6.0 & higher |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Brand Name: | GE DashPort 2 Patient Monitor |
| Model: | |
| Affected: | Serial Numbers containing "SE5" |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | The manufacturer had advised that there are several potential issues with the Dash 3000/4000/5000 patient monitors; 1/ Dash stops communication on the Unity Network resulting in loss of monitoring at the centra station 2/ Dash displays distorted ECG waveform in combination monitoring mode 3/ Dash incorrectly becomes Time Master on the Unity Network 4/ Dash date reverts to 1993 5/ DashPort 2 stops communicating on the Unity Network resulting in loss of monitoring at the central station |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/8/2012 |