Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13299
Brand Name:Covidien Cool-tip RFA Electrode Kits
Model:SWCT1530, SWCT15303, SWCT1540, SWCT2530, SWCT25303, SWCT2540
Affected:Various lots
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Brand Name:Covidien DGPHP High-Power single use grounding pads
Model:ACT1020, ACT1030, ACT1507, ACT1510, ACT1520, ACT1530, ACT2020, ACT2030, ACT2530, ACTC1025, ACTC1525, ACTC2025, DGPHP
Affected:Selected lots
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:The manufacturer advises there is potential for foil degradation on the DGP-HP RFA High-Power single use grounding pads. This foil degradation may result in an electrical disconnect and an undesirable thermal profile during use which may result in a patient burn at the pad site.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/8/2012
Hide menus