Detail of Recall

Type of Product:Device
Medsafe Reference:13288
Brand Name:Advanced Neuromodulation Systems Eon Mini
Model:65-3788
Affected:Selected serial numbers
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Advanced Neuromodulation Systems Inc.
Issue:The manufacturer had identified an issue relating to battery electrical short in devices which could have resulted in loss of communication with the device or the inability to recharge. When one of the battery terminals of the internal device battery came into contact with the circuit board, a short to the internal device battery could occur which may cause the IPG to not communicate or recharge, resulting in an explant.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Public Recall
Date Commenced:10/8/2012