Detail of Recall

Type of Product:Device
Medsafe Reference:13275
Brand Name:Siemens Advia Centaur HIV 1/0/2 Enhanced (EHIV) Assay
Model:01463908
Affected:Selected lots ending in 093, 095, 097, 101, 103, 105, 107, 109
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The Manufacturer confirms the ADVIA Centaur HIV 1/O/2 (EHIV) assay is not meeting its on-board stability claims of 28 days as stated in the IFU. Customers are asked to reduce the on-board stability of their EHIV assay reagents to 14 days and to reduce the calibration interval of their EHIV assay reagents to 7 days.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/8/2012