Type of Product:Device
Medsafe Reference:13263
Brand Name:Philips Essenta DR Compact X-Ray system
Model:Stand with detector carrier: 12NC
Affected:S/N 9890-010-88201
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:There is the potential for a serious injury if all of the following occurs:
1/ One of the Safety Springs breaks and
2/ U-arm is in the upper or middle position of its movement range when the spring breaks and
3/ Patient or operator is positioned below the U-arm.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/8/2012