Detail of Recall

Type of Product:Device
Medsafe Reference:13261
Brand Name:Philips EasyDiagnost X-Ray System
Model:Models 70642, 70643, 706030, 706031, 706032
Affected:Units that have had the inclination switch replaced during service
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:An inclination switch in the stand may have been mounted incorrectly during system service. As a result, the switch could not stop the tilting movement of the table base in the event of a combination of several hardware or software errors.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/8/2012