Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13261 |
| Brand Name: | Philips EasyDiagnost X-Ray System |
| Model: | Models 70642, 70643, 706030, 706031, 706032 |
| Affected: | Units that have had the inclination switch replaced during service |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | An inclination switch in the stand may have been mounted incorrectly during system service. As a result, the switch could not stop the tilting movement of the table base in the event of a combination of several hardware or software errors. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/8/2012 |