Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13242
Brand Name:Siemens Dimension Vista Theophylline reagent cartridge
Model:K4071
Affected:12135AE and 12171BB
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Manufacturer confirms that it has received complaints from customers of "Abnormal reaction" errors occurring on calibrations, QA and patient samples. The Abnormal Reaction flag is being triggered by an internal check in the method for atypical particle reagent aggregation. Manufacturer advises that results reported with the Abnormal reaction flag are acurate, however per the Vista Operator's Guide, flagged results should not be reported and should be repeated. This is considered no health risk.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/8/2012