Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13193 |
| Brand Name: | Abbott RealTime High Risk HPV Amplification Reagent Kit |
| Model: | 02N09-90 |
| Affected: | Lot 96539HN00 |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Molecular Inc |
| Issue: | The manufacturer has received reports of barcodes on reagent packs being unable to be scanned due to the barcodes being diffuse and/or having white spots within the barcode which prevents them from being read consistently by the m2000sp instrument. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/7/2012 |