Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13192 |
| Brand Name: | Philips Brilliance series 6, 16, 40, 64 CT system |
| Model: | |
| Affected: | Selected S/Ns |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Brand Name: | Philips CT Big Bore CT system |
| Model: | |
| Affected: | Selected S/Ns |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Brand Name: | Philips iCT and iCT SP CT systems |
| Model: | |
| Affected: | Selected S/Ns |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | In the CT systems above, sixteen bolts secure J-brackets which support the carbon table top to the couch subframe. Philips Healthcare received two complaints stating that the bolt heads of several of these sixteen bolts had broken off. A risk of serious injury (bone fracture, head injury, concussion) was identified in the event that all sixteen (16) bolts are broken. If this occurs, and the carbon table top is extended, the table top may tilt or tip over, or it may result in unintended pivotal motion of the table. If the patient is not secured to the table, there is a possibility the patient may slide off, which may cause serious injury. If less than sixteen (16) of the J-bracket bolts fail, the carbon table top will remain attached to the couch subframe. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/7/2012 |