Type of Product:Device
Medsafe Reference:13185
Brand Name:Philips MRx HeartStart Defibrillator
Model:models M3535A and M3536A, M3536J, M3536M, M3536MC, M3536M1, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6.
Affected:All units manufactured prior to June 2012
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:When HeartStart MRx Defibrillator Monitors are used in external transport and EMS environments the mechanical/electrical connection between the pads/paddles therapy cable (including pads CPR therapy cable and external paddles cable) and the MRx therapy connection port may experience higher than expected levels of stress causing accelerated wear. Without routine periodic inspections and preventive action by users, wear of the connection could ultimately prevent the device from sensing that the pads/paddles therapy cable is connected. This wear also may cause the MRx to inappropriately identify the pads therapy cable, external paddles, or internal paddles.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/7/2012