| Type of Product: | Device |
| Medsafe Reference: | 13154 |
| Brand Name: | Schiller FRED easy Defibrillators |
| Model: | All models |
| Affected: | All serial number |
| Software version: | |
| Recalling Organisation: | Schiller Australia Pty Ltd Unit 3/54 Pendlebury Road Cardiff NSW 2285 Australia |
| Contact Information: | |
| Manufacturer: | Schiller Medical SAS |
| Issue: | A false defibrillator error message, "transistor error", displays on screen and requires the device to be re-started. This problem may prevent shock delivery, which may result in failure to defibrillate and a reduced chance of survival for the patient. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Public Recall |
| Date Commenced: | 24/7/2012 |