Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13149 |
| Brand Name: | Sysmex CS-2100i/ 2000i Systems |
| Model: | |
| Affected: | all systems with software versions up to and including SW 00-60 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd 55 Hugo Johnston Drive Penrose Auckland |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Sysmex Corporation |
| Issue: | Manufacturer advises that CS-2000i and CS-2100i with software version 00-60 show the phenomenon that certain results with error flags highlighted for review, are automatically validated and sent to the host computer without those flags. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/7/2012 |