Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13148
Brand Name:ASP Sterrad 100S System Cassettes
Model:10113
Affected:12A070
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:Recall for a single lot of the Sterrad 100S System cassettes. Manufacturer has disccovered that printed information on the cassette results in the sterilizer being unable to properly read the barcode. During the automated process used to print information on Sterrad 100S System cassettes, an additional thirteenth (13th) bar was added to the barcode resulting in an unusable cassette.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/7/2012