Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13148 |
| Brand Name: | ASP Sterrad 100S System Cassettes |
| Model: | 10113 |
| Affected: | 12A070 |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | Recall for a single lot of the Sterrad 100S System cassettes. Manufacturer has disccovered that printed information on the cassette results in the sterilizer being unable to properly read the barcode. During the automated process used to print information on Sterrad 100S System cassettes, an additional thirteenth (13th) bar was added to the barcode resulting in an unusable cassette. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/7/2012 |