Detail of Recall

Type of Product:Device
Medsafe Reference:13130
Brand Name:Radiometer ABL90Flex Blood Gas analysers
Model:393-090
Affected:Software versions below V2.7 MR4
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:Radiometer Medical has become aware that there is a potential risk of reporting of current patient results to a previous patient ID. This may occur when an ABL90Flex analyzer with software versions below V2.7 MR4 is using the HL7 V2.5 communication protocol for transmitting results to a HIS/LIS system.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/7/2012