Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13130 |
| Brand Name: | Radiometer ABL90Flex Blood Gas analysers |
| Model: | 393-090 |
| Affected: | Software versions below V2.7 MR4 |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | Radiometer Medical has become aware that there is a potential risk of reporting of current patient results to a previous patient ID. This may occur when an ABL90Flex analyzer with software versions below V2.7 MR4 is using the HL7 V2.5 communication protocol for transmitting results to a HIS/LIS system. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/7/2012 |