Detail of Recall

Type of Product:Device
Medsafe Reference:13129
Brand Name:Ortho Clinical Vitros Chemistry products DGXN Slides
Model:8343386
Affected:All lots
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics c/- Johnson & Johnson
13A Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 800 705
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:The manufacturer has received complaints of elevated results using heparin plasma samples for VITROS Chemistry Products DGXN Slides. As a result of their investigation into these complaints, heparin plasma is being removed as a recommended specimen type for VITROS DGXN Slides.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/7/2012