Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13118
Brand Name:ASP Sterrad Cyclesure 24 Biological Indicator
Model:14324
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:An FDA review of ASP data showed that the Sterrad Cyclesure 24 Biological Indicators cannot effectively monitor the sterilization process throughout the indicated 15-month shelf life. After reviewing additional ASP data, the FDA believes that the product may be used with a 6-month shelf life.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/7/2012