Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13118 |
| Brand Name: | ASP Sterrad Cyclesure 24 Biological Indicator |
| Model: | 14324 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | An FDA review of ASP data showed that the Sterrad Cyclesure 24 Biological Indicators cannot effectively monitor the sterilization process throughout the indicated 15-month shelf life. After reviewing additional ASP data, the FDA believes that the product may be used with a 6-month shelf life. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/7/2012 |