Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13117 |
| Brand Name: | Elekta Monaco Radiation Treatment Planning System |
| Model: | |
| Affected: | Release 3.10 |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Brand Name: | Elekta Focal Radiation Therapy Treatment Planning System |
| Model: | |
| Affected: | Release 4.64.00 |
| Software version: | |
| Recalling Organisation: | AlphaXRT Ltd 10 The Promenade Takapuna Auckland |
| Contact Information: | |
| Manufacturer: | Elekta Inc |
| Issue: | If the DICOM coordinates option is turned on in the Reports pull down menu the following issues are seen on the resulting print out: Issue1: The coordinates will be in DICOM space, thus not matching the Setup Reference dialog. Issue 2: The shift coordinates are reported incorrectly. The Z coordinate should become DICOM negative Y and the Y coordinate should become DICOM Z. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/7/2012 |