Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13117
Brand Name:Elekta Monaco Radiation Treatment Planning System
Model:
Affected:Release 3.10
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Brand Name:Elekta Focal Radiation Therapy Treatment Planning System
Model:
Affected:Release 4.64.00
Software version:
Recalling Organisation:AlphaXRT Ltd
10 The Promenade
Takapuna
Auckland
Contact Information:
Manufacturer:Elekta Inc
Issue:If the DICOM coordinates option is turned on in the Reports pull down menu the following issues are seen on the resulting print out:
Issue1: The coordinates will be in DICOM space, thus not matching the Setup Reference dialog.
Issue 2: The shift coordinates are reported incorrectly. The Z coordinate should become DICOM negative Y and the Y coordinate should become DICOM Z.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/7/2012