Detail of Recall

Type of Product:Device
Medsafe Reference:13079
Brand Name:Invacare Birdie & Flamingo Lifter, Carabineer
Model:Birdie, Birdie Compact, Flamingo
Affected:Serial number SN 091105631721 and higher
Software version:
Recalling Organisation:Invacare NZ Ltd
4 Westfield Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 468 222
Manufacturer:Invacare Portugal
Issue:The carabineer used to attach the spreader bar to the lifter, may show some wear in case of intensive/ severe use. This carabineer needs to be checked and replaced if the thickness of its sections measures less than 6 mm. A change to the carabeineer's shape has also been made to reduce the potential for disengagement of the spreader bar.
Recall action type:Recall
Recall action:Product to be modified
Level of recall:Public Recall
Date Commenced:5/7/2012