Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:12921
Brand Name:PerkinEllmer LifeCycle for Prenatal Screening
Model:5002-0020
Affected:Software versions 3.0, 3.1, 3.2
Software version:
Recalling Organisation:SciMed Ltd
Unit 2, 1 Tussock Lane
Ferrymead
Christchurch 8023
Contact Information:0800-724-633
Manufacturer:PerkinElmer
Issue:Perkin Elmer has become aware about LifeCycle risk calculation problem that can potentially take place with assisted reproduction screening where an egg donor is involved. The change in calculated risk is dependent on the age difference between the mother and the egg donor. In the worst case, if the age of the mother is substantially higher than the egg donor's age. a true low risk value can be turned as high and thereby cause indirect harm because of possibility for an unnecessary medical intervention.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/12/2012