Detail of Recall

Type of Product:Device
Medsafe Reference:12808
Brand Name:Philips HeartStart MRx
Model:M3536A
Affected:Software versions, R.01.00, R.02.00 and R.03.01
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The affected devices may experience unexpected power cycling (system reboot) during device power on (startup). Some devices have been observed to experience rebooting cycles for up to five minutes before the MRx stabilizes. If the problem were to occur there is a potential that therapy may be delayed.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/2/2015